HPTLC for stability testing of dihydroartemisinin

    CBS Articles

    Authors: Akshay Charegaonkar, Tukaram Thite, Amit Palande, Dr. Saikat Mallick

    Published in CBS 130

Amit Palande, application specialist under the guidance of Akshay Charegaonkar (managing director), Tukaram Thite (senior lab manager) and Dr. Saikat Mallick (lab manager) work at Anchrom Enterprises Pvt Ltd, Mumbai, India. The company specializes in instrumental planar chromatography, and develops new, quantitative, and regulatory compliant analytical methods for pharmaceutical formulations, APIs, herbal products, food products, organic intermediates, dyes etc. Mr. Palande benefits from HPTLC because it is a fast, simple, economical and flexible, “visible chromatography” technique. HPTLC is risk-free and multiple detections can be made without repeating chromatography. This is especially helpful for dihydroartemisinin, because it is not UV absorbing and needs derivatization for densitometric detection.

Introduction

Dihydroartemisinin (also known as dihydroqinghaosu, artenimol or DHA) is a drug used to treat malaria. It is globally recognized for its efficacy and safety in the clinical treatment of malaria for decades. DHA is the active metabolite of all artemisinin compounds (artemisinin, artesunate, artemether, etc.) and is also available as a drug by itself. It is a semi-synthetic derivative of artemisinin and is widely used as an intermediate in the preparation of other artemisinin derived antimalarial drugs. DHA is often combined with piperaquine phosphate (PPQ). Like any formulation, these tablets need to be tested for shelf life i.e. stability.

Shelf-life studies are performed by accelerated or forced degradation studies as per ICH guidelines Q 1 A (R2).

Chemical structures of dihydroartemisinin and piperaquine phosphate

Chemical structures of dihydroartemisinin and piperaquine phosphate

HPTLC is well suited for stability studies, because it is inexpensive and time-saving. Accelerated degradation studies need a very large number of samples to be analyzed. By HPTLC, 15-20 samples can be quantified simultaneously on one plate in about 40–80 minutes. The solvent consumption is only about 20 mL for those 20 samples and little waste is produced. It was separately established, that piperaquine phosphate did not degrade in the studies. Hence a method was developed that kept piperaquine phosphate at the base of the chromatogram but selectively moved DHA and its two degradation products. DHA and the degradation products were detected by simple derivatization and then quantitatively evaluated.

HPTLC is a cost-effective and time-saving technique for the pharma industry, which deals with a heavy load of samples, also competition to introduce a new formulation is very intense. The presented method is a green method, which only uses 20 mL of solvent for 15–20 samples and produces almost no waste. Since the degradation products are unknown, the common procedure of establishing the calibration function using the diluted main substance (DHA in this case) was applied. Many times, as in this case, the samples have to be overloaded to detect the small quantities of impurities/degradation products.

Standard solutions

A stock solution of 1.0 mg/mL of dihydroartemisinin in acetonitrile is prepared and diluted 1:20 (V/V) for analysis.

Sample preparation

Ten tablets containing 40.0 mg of dihydroartemisinin and 320.0 mg of piperaquine phosphate are milled. Powdered tablets equivalent to 100.0 mg of DHA are transferred in a 20.0 mL volumetric flask and dissolved in acetonitrile. The supernatant after centrifugation is used for application.

Chromatogram layer

HPTLC plates silica gel 60 F254 (Merck), 20 x 10 cm are used.

Sample application

Samples and standard solutions are applied as bands with the Automatic TLC Sampler (ATS 4), 15 tracks, band length 8.0 mm, distance from left edge 20.0 mm, distance from lower edge 8.0 mm. 10 μL for sample and standard solutions are applied.

Chromatography

Plates are developed in a 20 x 10 cm Twin Trough Chamber with chamber saturation (with filter paper) for 20min, development with cyclohexane – ethyl acetate – glacial acetic acid 10:5:1 (V/V) to the migration distance of 90 mm (from the lower edge), followed by drying for 5 min.

Post-chromatographic derivatization

Plates are sprayed with 3.0 mL of anisaldehydesulfuric acid reagent (to 170 mL of cooled methanol, 20 mL of acetic acid and 10 mL of sulfuric acid are added, after cooling to room temperature 1mL of anisaldehyde is added) using the CAMAG Derivatizer. After spraying, the plates are heated at 110 °C for 5 min using the TLC Plate Heater.

Documentation

Images of the plate are captured with the TLC Visualizer in UV 366 nm and white light.

Densitometry

Absorbance measurement at 540 nm (tungsten lamp) is performed with CAMAG TLC Scanner 4 and visionCATS, slit dimension 6.00 mm x 0.45 mm, scanning speed 20 mm/s, evaluation via peak area.

Results and discussion

The determination of the degradation product with the mobile phase cyclohexane – ethyl acetate – glacial acetic acid 10:5:1 (V/V) is verified by the positions of the individual drugs, where PPQ remains at the application position and DHA moves to hRF 40. Any other zones at different hRF are reported as degradation products. The placebo shows no zones in the chromatogram. Prior to derivatization, no zones are visible. Quantitative evaluation is performed after derivatization by densitometric absorbance measurement at 540 nm. The data is recorded and used to calculate the degradation products. Total impurities in the samples are found to be within the 1.5% limit as per specification, defined in house. A representative densitogram of a stability batch sample and standard is shown. The area of each impurity is calculated as follows and then summarized.

Image in white light after derivatization

Image in white light after derivatization: Track 1: placebo, track 2: piperaquine phosphate, tracks 3, 4: DHA standard, tracks 5–10: different formulations of tablets in the stability study (DP = degradation product);

Densitogram at UV 540 nm of a tablet batch in which degradation products were detected.

Densitogram at UV 540 nm of a tablet batch in which degradation products were detected.
 

Results of a 6 months old batch

Results of a 6 months old batch
 

[1] CAMAG Application Note A-86.1: Determination of artemisinin in Artemisia annua leaf by HPTLC

Further information is available on request from the authors.