Product Release Information visionCATS 2.5 (2.5.18072.1)

Release Number: 2.5.18072.1
Release Date: 2018-02-22

Customer Support: For more information or support, please visit our website or email us at support@camag.com

Introduction

This document contains the release notes for visionCATS version 2.5. The following sections describe the release in detail and provide late-breaking or other information that supplements the main documentation.

Installation and Upgrade Notes

Installation

See the installation instructions for full details.

IMPORTANT: You must completely uninstall any previous "developer release" or "early access" version of this product before installing this release.

IMPORTANT: When upgrading from visionCATS 2.3 (with or without SP1), the installer will end with a message "Please (re)install .Net Framework 4.6.1 or greater." on machines having the .NET Framework 4.7 (either manually installed, via Windows Update or by default on Windows 10 Creators Update). This error can be safely ignored.

Manifest

This release consists of the following items:

Release notes
Drivers
Instrument Firmware
visionCATS installer
Example Files and Template Files (you may import them after installation

System Requirements

Recommended:
3rd gen (Ivy Bridge) Intel Core i5 (i5-3xxx) or equivalent
Nvidia GeForce GTX 760 or AMD Radeon R9 270
8 GB RAM
Windows 10 Pro (or Enterprise) 64-bit
True color 32bits, Full HD (or higher) monitor resolution with a Windows Aero theme enabled
1 TB Hard Drive (frequent user)
Mouse with a scroll wheel and 3 buttons

Minimum:
3rd gen (Ivy Bridge) Intel Core i3 (i3-3xxx) or equivalent
Nvidia GeForce GT 750M or Intel Iris Pro Graphics 580
4 GB RAM
Windows 7 Professional 64-bit
True color 32bits, 1280 x 1024 monitor resolution with a Windows Aero theme enabled
500 GB Hard Drive (occasional user)
Mouse or touchpad with 2 buttons

This version has been tested and released on Win7 64bit and Win10 64bit (see “Known issues” for detailed information).

For "server only" installation: This version has been tested and released on Windows Server 2012 R2 (domain controller excluded).

What's new in visionCATS 2.5?

  • Improved search in Explorer
  • New concentration unit type choices in evaluation
  • Analysis and Comparison Profile Wavelength view
  • Peak detection in Comparison Profile
  • HPTLC log in report
  • Open as
  • Notifications on analysis progress
  • Profile viewer with docking or multi windows
  • visionCATS Diagnostics

Known Problems and Workarounds

Issue: After re-installing with different server and instrument service ports, no instrument service are available.
WORKAROUND: Restart the service visionCATS Server.

Issue: The camera is not recognized anymore after the PC was in sleep or hibernate mode.
WORKAROUND: Some USB3 controllers have some issues with the camera, like the Intel® USB 3.0 eXtensible Host Controller. Restart the machine to resolve the problem, and disable sleep mode to prevent the problem from occurring again.

Issue: Blue screen when capture image with TLC Visualizer on a Windows 10
WORKAROUND: Go to Power Options/Choose what the power button does/Change settings that are currently available/Shutdown settings and disable the option “Turn on fast startup”

Issue: A click on the OK button of the message box appearing when an error occurs during a step execution, while the instrument is already executing a step of another analysis, causes an error to occur in the step currently in progress.
WORKAROUND: When the error message appears, click on the OK button before starting the execution of the other step.

Issue: The import of winCATS methods won't import steps of instruments not supported in visionCATS, and some winCATS specific parameters will be ignored.
REASON: winCATS and visionCATS don't exactly support the same set of instruments, and for each instrument, the exact same set of parameters
WORKAROUND: After having imported some winCATS methods, visionCATS will inform the user when some steps are ignored. In all cases, please review the imported methods.

Issue: After changing of the Visualizer, a Visualizer cannot be executed in an analysis already having images
REASON: Only one Visualizer can be used in an analysis
WORKAROUND: Redo the analysis

Issue: After changing of the camera for a Visualizer, an already opened analysis with a Visualizer step waiting for execution, or Visualizer Manual control or OQ may behave strangely (using the old camera, bad image corrections...)
WORKAROUND: Before changing of a camera, close all analyses with the next step as Visualizer and Visualizer views

Issue: After changing of the camera for a Visualizer, the Visualizer icon in the toolbox isn't marked as having a bad Diagnostic
WORKAROUND: Restart the client

Issue: While using two Visualizers on a single PC, one or both cameras aren't detected, or the camera of one Visualizer seems to be connected to the other.
WORKAROUND: We strongly suggest to use only one Visualizer per Instrument Service (PC)

Issue: No image appears in Data or in Comparison after the initial installation of visionCATS on the PC
WORKAROUND: Restart Windows

Issue: Sequence table doesn't allow to drag and drop a multi-selection
WORKAROUND: Before clicking for drag and during the drag, hold SHIFT key for dragging a multi-selection

Issue: Application crashes after clicking on an image or opening a comparison
WORKAROUND: Update your graphic driver to the latest version

Issue: After installation, no visionCATS is installed. Also, the SQL Server service may not want to start.
WORKAROUND: Change the name of the user which installs the application, so that the username is different from PC and domain name

Issue: Instruments aren't responding correctly
WORKAROUND: Check if you have an Intel(R) Active Management Technology - SOL COM port on your machine, and disable the corresponding COM port in visionCATS/General Settings

Issue: When using offline help, the display of some pages is very slow
REASON: the page requests an online resource that is not accessible, causing the slow down
WORKAROUND: use the online help when possible. If not possible, try using Chrome or Firefox (instead of Internet Explorer), where the slow down occurs less often.

Issue: visionCATS fails in Scanner/Scanner 3 Diagnostics if the Positioning-Test is started in the same session.
REASON: Scanner data of the new test cannot be stored for later processing. This does not invalidate the test results of the previous run.
WORKAROUND: Close the current Scanner Instrument Tab and reopen it. Attention: just switching between the Tabs inside the Instrument Tab will not prevent the failure. You need to close the Instrument Tab.

Issue: When installing visionCATS' client on machines with Windows of different language (different code page), or restoring a backup on a machine with Windows of another language (code page), all passwords are corrupted.
REASON: Passwords are encrypted using the code page of the local machine (client).
WORKAROUND: All visionCATS machines of a system (clients and server) should have the same Windows code page.